Abstract
Axial-flow ventricular assist devices are being increasingly used to support hemodynamically compromised patients undergoing percutaneous coronary intervention. Periprocedural valvular complications have been recognized in a few case reports. We present a unique case of entanglement of the Impella within the mitral subvalvular apparatus, retrieved successfully using a snare under fluoroscopic guidance. (Level of Difficulty: Advanced.)
Citation
Elsherif A, Nadir A, Ludman PF, Khan SQ. Retrieval of Entrapped Catheter-Mounted Axial Flow Pump From Mitral Subvalvular Apparatus Using a Snare Catheter. JACC Case Rep. 2021 Oct 6;3(13):1494-1498. doi: 10.1016/j.jaccas.2021.06.041.