Publication

Estimation of infant radiation exposure from ingestion of breastmilk after SeHCAT capsule administration

Rowley, Lisa
Citations
Altmetric:
Affiliation
Comber, E.A., Adesanya, O., Gladstone, J., Coleclough, S., Rowley, L.,
Other Contributors
Publication date
2023-01-31
Subject
Research Projects
Organizational Units
Journal Issue
Abstract
im: Assessment of the imaging properties of 3D-printable materials using dual energy computed tomography (DECT) to match clinical values for imaging phantoms. Methods: 3D-printed samples were imaged using DECT. Regions of interest were analyzed to assess spectral computed tomography (CT) numbers at various energies and measure the electron density (ρe) and effective atomic number (Zeff). Results: Electron density was proportional to the CT number for the materials assessed with Zeff between 6.43 and 7.01. The measured CT number increased with monochromatic energy for all but one sample. Conclusion: A single DECT scan provides valuable information regarding the properties of 3D-printable material due to the ease of measurement of ρe and Zeff. The majority of 3D-printed materials analyzed behaved like adipose tissue across a range of energies in CT imaging.
Citation
Rowley,L.M., Comber, E.
Type
Article
Description
Additional Links
DOI
PMID
Journal
Embedded videos